Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 201
:CHANnel<n>:PROBe:SKEW
(see page 658)
Command Syntax
:CHANnel<n>:PROBe:SKEW <skew value>
<skew value> ::= skew time in NR3 format
<skew value> ::= -100 ns to +100 ns
<n>::={1|2|3|4}
The :CHANnel<n>:PROBe:SKEW command sets the channel- to- channel
skew factor for the specified channel. Each analog channel can be adjusted
+ or - 100 ns for a total of 200 ns difference between channels. You can
use the oscilloscope's probe skew control to remove cable-delay errors
between channels.
Query Syntax
:CHANnel<n>:PROBe:SKEW?
The :CHANnel<n>:PROBe:SKEW? query returns the current probe skew
setting for the selected channel.
Return Format
<skew value><NL>
<skew value> ::= skew value in NR3 format
See Also "Introduction to :CHANnel<n> Commands" on page 190

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals