Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 225
:EXTernal:PROTection
(see page 658)
Command Syntax
:EXTernal:PROTection[:CLEar]
When the external trigger input impedance is set to 50, the external
trigger input is protected against overvoltage. When an overvoltage
condition is sensed, the input impedance for the external trigger is
automatically changed to 1 M
. The :EXTernal:PROTection[:CLEar]
command is used to clear (reset) the overload protection. It allows the
external trigger to be used again in 50
mode after the signal that caused
the overload has been removed from the external trigger input. Reset the
external trigger input impedance to 50
(see ":EXTernal:IMPedance" on
page 221) after clearing the overvoltage protection.
Query Syntax
:EXTernal:PROTection?
The :EXTernal:PROTection query returns the state of the input protection
for external trigger. If the external trigger input has experienced an
overload, TRIP (tripped) will be returned; otherwise NORM (normal) is
returned.
Return Format
{NORM | TRIP}<NL>
See Also "Introduction to :EXTernal Trigger Commands" on page 218
":EXTernal:IMPedance" on page 221
":EXTernal:PROBe" on page 222

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals