Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 259
:MARKer:X1Position
(see page 658)
Command Syntax
:MARKer:X1Position <position> [suffix]
<position> ::= X1 cursor position in NR3 format
<suffix> ::= {s | ms | us | ns | ps | Hz | kHz | MHz}
The :MARKer:X1Position command:
• Sets :MARKer:MODE to MANual if it is not currently set to WAVeform
(see ":MARKer:MODE" on page 258).
• Sets the X1 cursor position to the specified value.
Query Syntax
:MARKer:X1Position?
The :MARKer:X1Position? query returns the current X1 cursor position.
This is functionally equivalent to the obsolete :MEASure:TSTArt
command/query.
Return Format
<position><NL>
<position> ::= X1 cursor position in NR3 format
See Also • "Introduction to :MARKer Commands" on page 257
• ":MARKer:MODE" on page 258
• ":MARKer:X2Position" on page 261
• ":MARKer:X1Y1source" on page 260
• ":MARKer:X2Y2source" on page 262
• ":MEASure:TSTArt" on page 591
NOTE
If the front-panel cursors are off, the marker position values are not defined and an error is
generated. Make sure to set :MARKer:MODE to MANual or WAVeform to put the cursors in
the front-panel Normal mode.