Page #587 background image
Obsolete and Discontinued Commands 7
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 587
:MEASure:TDELta
(see page 658)
Query Syntax
:MEASure:TDELta?
The :MEASure:TDELta? query returns the time difference between the
Tstop marker (X2 cursor) and the Tstart marker (X1 cursor).
Tdelta = Tstop - Tstart
Tstart is the time at the start marker (X1 cursor) and Tstop is the time at
the stop marker (X2 cursor). No measurement is made when the
:MEASure:TDELta? query is received by the oscilloscope. The delta time
value that is output is the current value. This is the same value as the
front-panel cursors delta X value.
Return Format
<value><NL>
<value> ::= time difference between start and stop markers in NR3 format
See Also "Introduction to :MARKer Commands" on page 257
"Introduction to :MEASure Commands" on page 272
":MARKer:X1Position" on page 259
":MARKer:X2Position" on page 261
":MARKer:XDELta" on page 263
":MEASure:TSTArt" on page 591
":MEASure:TSTOp" on page 592
NOTE
The :MEASure:TDELta command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:XDELta command (see page 263) instead.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals