478 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:SPI:SOURce:FRAMe
(see page 658)
Command Syntax
:TRIGger:SPI:SOURce:FRAMe <source>
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital0,..,DIGital15} for the MSO models
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger:SPI:SOURce:FRAMe command sets the frame source when
:TRIGger:SPI:FRAMing is set to CHIPselect or NOTChipselect.
Query Syntax
:TRIGger:SPI:SOURce:FRAMe?
The :TRIGger:SPI:SOURce:FRAMe? query returns the current frame source
for the SPI serial frame.
Return Format
<source><NL>
See Also "Introduction to :TRIGger Commands" on page 411
":TRIGger:SPI:SOURce:CLOCk" on page 476
":TRIGger:SPI:SOURce:DATA" on page 477
":TRIGger:SPI:FRAMing" on page 473