472 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:SPI:CLOCk:TIMeout
(see page 658)
Command Syntax
:TRIGger:SPI:CLOCk:TIMeout <time_value>
<time_value> ::= time in seconds in NR1 format
The :TRIGger:SPI:CLOCk:TIMeout command sets the SPI signal clock
timeout resource in seconds from 500 ns to 10 s when the
:TRIGger:SPI:FRAMing command is set to TIMeout. The timer is used to
frame a signal by a clock timeout.
Query Syntax
:TRIGger:SPI:CLOCk:TIMeout?
The :TRIGger:SPI:CLOCk:TIMeout? query returns current SPI clock timeout
setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR1 format
See Also "Introduction to :TRIGger Commands" on page 411
":TRIGger:SPI:CLOCk:SLOPe" on page 471
":TRIGger:SPI:SOURce:CLOCk" on page 476
":TRIGger:SPI:FRAMing" on page 473

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals