Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 405
:TIMebase:REFerence
(see page 658)
Command Syntax
:TIMebase:REFerence <reference>
<reference> ::= {LEFT | CENTer | RIGHt}
The :TIMebase:REFerence command sets the time reference to one division
from the left side of the screen, to the center of the screen, or to one
division from the right side of the screen. Time reference is the point on
the display where the trigger point is referenced.
Query Syntax
:TIMebase:REFerence?
The :TIMebase:REFerence? query returns the current display reference for
the main window.
Return Format
<reference><NL>
<reference> ::= {LEFT | CENT | RIGH}
See Also "Introduction to :TIMebase Commands" on page 400
":TIMebase:MODE" on page 402
Example Code
' TIME_REFERENCE - Possible values are LEFT and CENTER.
' - LEFT sets the display reference on time division from the left.
' - CENTER sets the display reference to the center of the screen.
myScope.WriteString ":TIMEBASE:REFERENCE CENTER" ' Set reference to
center.
Example program from the start: "VISA COM Example in Visual Basic" on
page 744

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals