380 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:SBUS:DISPlay
(see page 658)
Command Syntax
:SBUS:DISPlay <display>
<display> ::= {{1 | ON} | {0 | OFF}}
The :SBUS:DISPlay command turns displaying of the serial decode bus on
or off.
Query Syntax
:SBUS:DISPlay?
The :SBUS:DISPlay? query returns the current display setting of the serial
decode bus.
Return Format
<display><NL>
<display> ::= {0 | 1}
Errors "- 241, Hardware missing" on page 617
See Also "Introduction to :SBUS Commands" on page 374
":CHANnel<n>:DISPlay" on page 194
":VIEW" on page 162
":BLANk" on page 130
":STATus" on page 159
NOTE
This command is only valid on 4 (analog) channel oscilloscope models when a serial decode
option has been licensed.