416 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:HOLDoff
(see page 658)
Command Syntax
:TRIGger:HOLDoff <holdoff_time>
<holdoff_time> ::= 60 ns to 10 s in NR3 format
The :TRIGger:HOLDoff command defines the holdoff time value in seconds.
Holdoff keeps a trigger from occurring until after a certain amount of time
has passed since the last trigger. This feature is valuable when a waveform
crosses the trigger level multiple times during one period of the waveform.
Without holdoff, the oscilloscope could trigger on each of the crossings,
producing a confusing waveform. With holdoff set correctly, the
oscilloscope always triggers on the same crossing. The correct holdoff
setting is typically slightly less than one period.
Query Syntax
:TRIGger:HOLDoff?
The :TRIGger:HOLDoff? query returns the holdoff time value for the
current trigger mode.
Return Format
<holdoff_time><NL>
<holdoff_time> ::= the holdoff time value in seconds in NR3 format.
See Also "Introduction to :TRIGger Commands" on page 411

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals