442 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger[:EDGE]:REJect
(see page 658)
Command Syntax
:TRIGger[:EDGE]:REJect <reject>
<reject> ::= {OFF | LFReject | HFReject}
The :TRIGger[:EDGE]:REJect command turns the low- frequency or
high- frequency reject filter on or off. You can turn on one of these filters
at a time.
The high frequency reject filter adds a 50 kHz low-pass filter in the
trigger path to remove high frequency components from the trigger
waveform. Use the high frequency reject filter to remove high- frequency
noise, such as AM or FM broadcast stations, from the trigger path.
The low frequency reject filter adds a 50 kHz high- pass filter in series
with the trigger waveform to remove any unwanted low frequency
components from a trigger waveform, such as power line frequencies,
that can interfere with proper triggering.
Query Syntax
:TRIGger[:EDGE]:REJect?
The :TRIGger[:EDGE]:REJect? query returns the current status of the reject
filter.
Return Format
<reject><NL>
<reject> ::= {OFF | LFR | HFR}
See Also "Introduction to :TRIGger Commands" on page 411
":TRIGger:HFReject" on page 415
":TRIGger[:EDGE]:COUPling" on page 440
NOTE
The :TRIGger[:EDGE]:REJect and the :TRIGger[:EDGE]:COUPling selections are coupled.
Changing the setting of the :TRIGger[:EDGE]:COUPling can change the COUPling setting.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals