580 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
7 Obsolete and Discontinued Commands
:HARDcopy:FILename
(see page 658)
Command Syntax
:HARDcopy:FILename <string>
<string> ::= quoted ASCII string
The HARDcopy:FILename command sets the output filename for those
print formats whose output is a file.
Query Syntax
:HARDcopy:FILename?
The :HARDcopy:FILename? query returns the current hardcopy output
filename.
Return Format
<string><NL>
<string> ::= quoted ASCII string
See Also "Introduction to :HARDcopy Commands" on page 246
":HARDcopy:FORMat" on page 581
NOTE
The :HARDcopy:FILename command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :SAVE:FILename command (see page 359) and
:RECall:FILename command (see page 352) instead.