Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 443
:TRIGger[:EDGE]:SLOPe
(see page 658)
Command Syntax
:TRIGger[:EDGE]:SLOPe <slope>
<slope> ::= {NEGative | POSitive | ALTernate}
The :TRIGger[:EDGE]:SLOPe command specifies the slope of the edge for
the trigger. The SLOPe command is not valid in TV trigger mode. Instead,
use :TRIGger:TV:POLarity to set the polarity in TV trigger mode.
Query Syntax
:TRIGger[:EDGE]:SLOPe?
The :TRIGger[:EDGE]:SLOPe? query returns the current trigger slope.
Return Format
<slope><NL>
<slope> ::= {NEG | POS | ALT}
See Also "Introduction to :TRIGger Commands" on page 411
":TRIGger:MODE" on page 417
":TRIGger:TV:POLarity" on page 482
Example Code
' TRIGGER_EDGE_SLOPE - Sets the slope of the edge for the trigger.
' Set the slope to positive.
myScope.WriteString ":TRIGGER:EDGE:SLOPE POSITIVE"
Example program from the start: "VISA COM Example in Visual Basic" on
page 744

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals