EasyManua.ls Logo

Agilent Technologies InfiniiVision 5000 Series

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
238 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:FUNCtion:RANGe
(see page 658)
Command Syntax
:FUNCtion:RANGe <range>
<range> ::= the full-scale vertical axis value in NR3 format.
The :FUNCtion:RANGe command defines the full- scale vertical axis for the
selected function.
Query Syntax
:FUNCtion:RANGe?
The :FUNCtion:RANGe? query returns the current full- scale range value for
the selected function.
Return Format
<range><NL>
<range> ::= the full-scale vertical axis value in NR3 format.
The range for ADD, SUBT, MULT is 8E-6 to 800E+3.
The range for the INTegrate function is 8E- 9 to 400E+3 (depends on sweep
speed).
The range for the DIFFerentiate function is 80E- 3 to 8.0E12 (depends on
sweep speed).
The range for the FFT (Fast Fourier Transform) function is 8 to 800 dBV.
See Also "Introduction to :FUNCtion Commands" on page 230
":FUNCtion:SCALe" on page 240

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Related product manuals