Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 437
:TRIGger:DURation:QUALifier
(see page 658)
Command Syntax
:TRIGger:DURation:QUALifier <qualifier>
<qualifier> ::= {GREaterthan | LESSthan | INRange | OUTRange | TIMeout}
The :TRIGger:DURation:QUALifier command qualifies the trigger duration.
Set the GREaterthan qualifier value with the
:TRIGger:DURation:GREaterthan command.
Set the LESSthan qualifier value with the :TRIGger:DURation:LESSthan
command.
Set the INRange and OUTRange qualifier values with the
:TRIGger:DURation:RANGe command.
Set the TIMeout qualifier value with the :TRIGger:DURation:GREaterthan
command.
Query Syntax
:TRIGger:DURation:QUALifier?
The :TRIGger:DURation:QUALifier? query returns the trigger duration
qualifier.
Return Format
<qualifier><NL>
See Also "Introduction to :TRIGger Commands" on page 411
":TRIGger:DURation:GREaterthan" on page 434
":TRIGger:DURation:LESSthan" on page 435
":TRIGger:DURation:RANGe" on page 438

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals