172 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:ACQuire:SEGMented:COUNt
(see page 658)
Command Syntax
:ACQuire:SEGMented:COUNt <count>
<count> ::= an integer from 2 to 250 in NR1 format
The :ACQuire:SEGMented:COUNt command sets the number of memory
segments to acquire.
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command, and data is acquired using the :DIGitize,
:SINGle, or :RUN commands. The number of memory segments in the
current acquisition is returned by the :WAVeform:SEGMented:COUNt?
query.
Query Syntax
:ACQuire:SEGMented:COUNt?
The :ACQuire:SEGMented:COUNt? query returns the current count setting.
Return Format
<count><NL>
<count> ::= an integer from 2 to 250 in NR1 format
See Also ":ACQuire:MODE" on page 169
":DIGitize" on page 132
":SINGle" on page 158
":RUN" on page 156
":WAVeform:SEGMented:COUNt" on page 519
":ACQuire:SEGMented:ANALyze" on page 171
"Introduction to :ACQuire Commands" on page 163
Example Code "Example Code" on page 173
NOTE
This command is available when the segmented memory option (Option SGM) is enabled.