Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 409
:TIMebase:WINDow:RANGe
(see page 658)
Command Syntax
:TIMebase:WINDow:RANGe <range value>
<range value> ::= range value in seconds in NR3 format
The :TIMebase:WINDow:RANGe command sets the full- scale horizontal time
in seconds for the zoomed (delayed) window. The range is 10 times the
current zoomed view window seconds per division setting. The main sweep
range determines the range for this command. The maximum value is one
half of the :TIMebase:RANGe value.
Query Syntax
:TIMebase:WINDow:RANGe?
The :TIMebase:WINDow:RANGe? query returns the current window
timebase range setting.
Return Format
<value><NL>
<value> ::= range value in seconds
See Also "Introduction to :TIMebase Commands" on page 400
":TIMebase:RANGe" on page 404
":TIMebase:POSition" on page 403
":TIMebase:SCALe" on page 406

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals