EasyManuals Logo

Agilent Technologies InfiniiVision 5000 Series Programmer's Guide

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #339 background imageLoading...
Page #339 background image
Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 339
:MTESt:RMODe:FACTion:SAVE
(see page 658)
Command Syntax
:MTESt:RMODe:FACTion:SAVE <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:SAVE command sets saving on mask failures
on or off.
See ":SAVE Commands" on page 357 for more information on save options.
Query Syntax
:MTESt:RMODe:FACTion:SAVE?
The :MTESt:RMODe:FACTion:SAVE? query returns the current mask failure
save setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :MTESt Commands" on page 320
":MTESt:RMODe:FACTion:PRINt" on page 338
":MTESt:RMODe:FACTion:STOP" on page 340
NOTE
Setting :MTESt:RMODe:FACTion:SAVE ON automatically sets
:MTESt:RMODe:FACTion:PRINt OFF.