Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 465
:TRIGger:LIN:SIGNal:BAUDrate
(see page 658)
Command Syntax
:TRIGger:LIN:SIGNal:BAUDrate <baudrate>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments
The :TRIGger:LIN:SIGNal:BAUDrate command sets the standard baud rate
of the LIN signal from 2400 b/s to 625 kb/s in 100 b/s increments. If you
enter a baud rate that is not divisible by 100 b/s, the baud rate is set to
the nearest baud rate divisible by 100 b/s.
Query Syntax
:TRIGger:LIN:SIGNal:BAUDrate?
The :TRIGger:LIN:SIGNal:BAUDrate? query returns the current LIN baud
rate setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments
See Also "Introduction to :TRIGger Commands" on page 411
":TRIGger:MODE" on page 417
":TRIGger:LIN:TRIGger" on page 469
":TRIGger:LIN:SIGNal:DEFinition" on page 612
":TRIGger:LIN:SOURce" on page 466

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals