EasyManuals Logo

Agilent Technologies InfiniiVision 5000 Series Programmer's Guide

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #519 background imageLoading...
Page #519 background image
Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 519
:WAVeform:SEGMented:COUNt
(see page 658)
Query Syntax
:WAVeform:SEGMented:COUNt?
The :WAVeform:SEGMented:COUNt query returns the number of memory
segments in the acquired data. You can use the
:WAVeform:SEGMented:COUNt? query while segments are being acquired
(although :DIGitize blocks subsequent queries until the full segmented
acquisition is complete).
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command. The number of segments to acquire is set using
the :ACQuire:SEGMented:COUNt command, and data is acquired using the
:DIGitize, :SINGle, or :RUN commands.
Return Format
<count> ::= an integer from 2 to 250 in NR1 format (count set by
:ACQuire:SEGMented:COUNt).
See Also ":ACQuire:MODE" on page 169
":ACQuire:SEGMented:COUNt" on page 172
":DIGitize" on page 132
":SINGle" on page 158
":RUN" on page 156
"Introduction to :WAVeform Commands" on page 502
Example Code "Example Code" on page 173
NOTE
This command is available when the segmented memory option (Option SGM) is enabled.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals