Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 519
:WAVeform:SEGMented:COUNt
(see page 658)
Query Syntax
:WAVeform:SEGMented:COUNt?
The :WAVeform:SEGMented:COUNt query returns the number of memory
segments in the acquired data. You can use the
:WAVeform:SEGMented:COUNt? query while segments are being acquired
(although :DIGitize blocks subsequent queries until the full segmented
acquisition is complete).
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command. The number of segments to acquire is set using
the :ACQuire:SEGMented:COUNt command, and data is acquired using the
:DIGitize, :SINGle, or :RUN commands.
Return Format
<count> ::= an integer from 2 to 250 in NR1 format (count set by
:ACQuire:SEGMented:COUNt).
See Also ":ACQuire:MODE" on page 169
":ACQuire:SEGMented:COUNt" on page 172
":DIGitize" on page 132
":SINGle" on page 158
":RUN" on page 156
"Introduction to :WAVeform Commands" on page 502
Example Code "Example Code" on page 173
NOTE
This command is available when the segmented memory option (Option SGM) is enabled.