EasyManua.ls Logo

Agilent Technologies InfiniiVision 5000 Series

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 195
:CHANnel<n>:IMPedance
(see page 658)
Command Syntax
:CHANnel<n>:IMPedance <impedance>
<impedance> ::= {ONEMeg | FIFTy}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:IMPedance command selects the input impedance setting
for the specified analog channel. The legal values for this command are
ONEMeg (1 M
) and FIFTy (50).
Query Syntax
:CHANnel<n>:IMPedance?
The :CHANnel<n>:IMPedance? query returns the current input impedance
setting for the specified channel.
Return Format
<impedance value><NL>
<impedance value> ::= {ONEM | FIFT}
See Also "Introduction to :CHANnel<n> Commands" on page 190

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Related product manuals