EasyManuals Logo

Agilent Technologies InfiniiVision 5000 Series Programmer's Guide

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #454 background imageLoading...
Page #454 background image
454 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:IIC:PATTern:ADDRess
(see page 658)
Command Syntax
:TRIGger:IIC:PATTern:ADDRess <value>
<value> ::= integer or <string>
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F}
The :TRIGger:IIC:PATTern:ADDRess command sets the address for IIC
data.The address can range from 0x00 to 0x7F (7- bit) or 0x3FF (10- bit)
hexadecimal. Use the don't care address (- 1 or 0xFFFFFFFF) to ignore the
address value.
Query Syntax
:TRIGger:IIC:PATTern:ADDRess?
The :TRIGger:IIC:PATTern:ADDRess? query returns the current address for
IIC data.
Return Format
<value><NL>
<value> ::= integer
See Also "Introduction to :TRIGger Commands" on page 411
":TRIGger:IIC:PATTern:DATA" on page 455
":TRIGger:IIC:PATTern:DATa2" on page 456
":TRIGger:IIC:TRIGger[:TYPE]" on page 460

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals