454 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:IIC:PATTern:ADDRess
(see page 658)
Command Syntax
:TRIGger:IIC:PATTern:ADDRess <value>
<value> ::= integer or <string>
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F}
The :TRIGger:IIC:PATTern:ADDRess command sets the address for IIC
data.The address can range from 0x00 to 0x7F (7- bit) or 0x3FF (10- bit)
hexadecimal. Use the don't care address (- 1 or 0xFFFFFFFF) to ignore the
address value.
Query Syntax
:TRIGger:IIC:PATTern:ADDRess?
The :TRIGger:IIC:PATTern:ADDRess? query returns the current address for
IIC data.
Return Format
<value><NL>
<value> ::= integer
See Also "Introduction to :TRIGger Commands" on page 411
":TRIGger:IIC:PATTern:DATA" on page 455
":TRIGger:IIC:PATTern:DATa2" on page 456
":TRIGger:IIC:TRIGger[:TYPE]" on page 460