Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 351
:RECall Commands
Recall previously saved oscilloscope setups and traces. See "Introduction to
:RECall Commands" on page 351.
Introduction to
:RECall
Commands
The :RECall subsystem provides commands to recall previously saved
oscilloscope setups and traces.
Reporting the Setup
Use :RECall? to query setup information for the RECall subsystem.
Return Format
The following is a sample response from the :RECall? query. In this case,
the query was issued following the *RST command.
:REC:FIL "scope_0"
Table 55 :RECall Commands Summary
Command Query Options and Query Returns
:RECall:FILename
<base_name> (see
page 352)
:RECall:FILename?
(see page 352)
<base_name> ::= quoted ASCII
string
:RECall:IMAGe[:STARt]
[<file_spec>] (see
page 353)
n/a <file_spec> ::= {<internal_loc>
| <file_name>}
<internal_loc> ::= 0-9; an
integer in NR1 format
<file_name> ::= quoted ASCII
string
:RECall:MASK[:STARt]
[<file_spec>] (see
page 354)
n/a <file_spec> ::= {<internal_loc>
| <file_name>}
<internal_loc> ::= 0-3; an
integer in NR1 format
<file_name> ::= quoted ASCII
string
:RECall:PWD
<path_name> (see
page 355)
:RECall:PWD? (see
page 355)
<path_name> ::= quoted ASCII
string
:RECall:SETup[:STARt]
[<file_spec>] (see
page 356)
n/a <file_spec> ::= {<internal_loc>
| <file_name>}
<internal_loc> ::= 0-9; an
integer in NR1 format
<file_name> ::= quoted ASCII
string