128 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:AUToscale:AMODE
(see page 658)
Command Syntax
:AUToscale:AMODE <value>
<value> ::= {NORMal | CURRent}
The :AUTOscale:AMODE command specifies the acquisition mode that is
set by subsequent :AUToscales.
When NORMal is selected, an :AUToscale command sets the NORMal
acquisition type and the RTIMe (real- time) acquisition mode.
When CURRent is selected, the current acquisition type and mode are
kept on subsequent :AUToscales.
Use the :ACQuire:TYPE and :ACQuire:MODE commands to set the
acquisition type and mode.
Query Syntax
:AUToscale:AMODE?
The :AUToscale:AMODE? query returns the autoscale acquire mode setting.
Return Format
<value><NL>
<value> ::= {NORM | CURR}
See Also "Introduction to Root (:) Commands" on page 124
":AUToscale" on page 126
":AUToscale:CHANnels" on page 129
":ACQuire:TYPE" on page 177
":ACQuire:MODE" on page 169