Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 203
:CHANnel<n>:PROTection
(see page 658)
Command Syntax
:CHANnel<n>:PROTection[:CLEar]
<n>::={1|2|3|4}
When the analog channel input impedance is set to 50, the input
channels are protected against overvoltage. When an overvoltage condition
is sensed, the input impedance for the channel is automatically changed to
1 M
. The :CHANnel<n>:PROTection[:CLEar] command is used to clear
(reset) the overload protection. It allows the channel to be used again in
50
mode after the signal that caused the overload has been removed
from the channel input. Reset the analog channel input impedance to 50
(see ":CHANnel<n>:IMPedance" on page 195) after clearing the overvoltage
protection.
Query Syntax
:CHANnel<n>:PROTection?
The :CHANnel<n>:PROTection query returns the state of the input
protection for CHANnel<n>. If a channel input has experienced an
overload, TRIP (tripped) will be returned; otherwise NORM (normal) is
returned.
Return Format
{NORM | TRIP}<NL>
See Also "Introduction to :CHANnel<n> Commands" on page 190
":CHANnel<n>:COUPling" on page 193
":CHANnel<n>:IMPedance" on page 195
":CHANnel<n>:PROBe" on page 199

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals