202 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:CHANnel<n>:PROBe:STYPe
(see page 658)
Command Syntax
:CHANnel<n>:PROBe:STYPe <signal type>
<signal type> ::= {DIFFerential | SINGle}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:PROBe:STYPe command sets the channel probe signal
type (STYPe) to differential or single- ended when using the 113xA Series
probes and determines how offset is applied.
When single- ended is selected, the :CHANnel<n>:OFFset command changes
the offset value of the probe amplifier. When differential is selected, the
:CHANnel<n>:OFFset command changes the offset value of the channel
amplifier.
Query Syntax
:CHANnel<n>:PROBe:STYPe?
The :CHANnel<n>:PROBe:STYPe? query returns the current probe signal
type setting for the selected channel.
Return Format
<signal type><NL>
<signal type> ::= {DIFF | SING}
See Also "Introduction to :CHANnel<n> Commands" on page 190
":CHANnel<n>:OFFSet" on page 198
NOTE
This command is valid only for the 113xA Series probes.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals