EasyManuals Logo

Agilent Technologies InfiniiVision 5000 Series Programmer's Guide

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #236 background imageLoading...
Page #236 background image
236 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:FUNCtion:OFFSet
(see page 658)
Command Syntax
:FUNCtion:OFFSet <offset>
<offset> ::= the value at center screen in NR3 format.
The :FUNCtion:OFFSet command sets the voltage or vertical value
represented at center screen for the selected function. The range of legal
values is generally +/- 10 times the current scale of the selected function,
but will vary by function. If you set the offset to a value outside of the
legal range, the offset value is automatically set to the nearest legal value.
Query Syntax
:FUNCtion:OFFSet?
The :FUNCtion:OFFSet? query outputs the current offset value for the
selected function.
Return Format
<offset><NL>
<offset> ::= the value at center screen in NR3 format.
See Also "Introduction to :FUNCtion Commands" on page 230
":FUNCtion:RANGe" on page 238
":FUNCtion:REFerence" on page 239
":FUNCtion:SCALe" on page 240
NOTE
The :FUNCtion:OFFset command is equivalent to the :FUNCtion:REFerence command.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals