Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 445
:TRIGger:GLITch Commands
Table 64 :TRIGger:GLITch Commands Summary
Command Query Options and Query Returns
:TRIGger:GLITch:GREat
erthan
<greater_than_time>[s
uffix] (see page 446)
:TRIGger:GLITch:GREat
erthan? (see
page 446)
<greater_than_time> ::=
floating-point number in NR3
format
[suffix] ::= {s | ms | us | ns |
ps}
:TRIGger:GLITch:LESSt
han
<less_than_time>[suff
ix] (see page 447)
:TRIGger:GLITch:LESSt
han? (see page 447)
<less_than_time> ::=
floating-point number in NR3
format
[suffix] ::= {s | ms | us | ns |
ps}
:TRIGger:GLITch:LEVel
<level> [<source>]
(see page 448)
:TRIGger:GLITch:LEVel
? (see page 448)
For internal triggers, <level>
::= .75 x full-scale voltage from
center screen in NR3 format.
For external triggers, <level>
::= ±(external range setting) in
NR3 format.
<source> ::= {CHANnel<n> |
EXTernal}
<n> ::= 1-2 or 1-4 in NR1 format
:TRIGger:GLITch:POLar
ity <polarity> (see
page 449)
:TRIGger:GLITch:POLar
ity? (see page 449)
<polarity> ::= {POSitive |
NEGative}
:TRIGger:GLITch:QUALi
fier <qualifier> (see
page 450)
:TRIGger:GLITch:QUALi
fier? (see page 450)
<qualifier> ::= {GREaterthan |
LESSthan | RANGe}
:TRIGger:GLITch:RANGe
<less_than_time>[suff
ix],
<greater_than_time>[s
uffix] (see page 451)
:TRIGger:GLITch:RANGe
? (see page 451)
<less_than_time> ::= 15 ns to
10 seconds in NR3 format
<greater_than_time> ::= 10 ns to
9.99 seconds in NR3 format
[suffix] ::= {s | ms | us | ns |
ps}
:TRIGger:GLITch:SOURc
e <source> (see
page 452)
:TRIGger:GLITch:SOURc
e? (see page 452)
<source> ::= {CHANnel<n> |
EXTernal}
<n> ::= 1-2 or 1-4 in NR1 format