Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 341
:MTESt:RMODe:SIGMa
(see page 658)
Command Syntax
:MTESt:RMODe:SIGMa <level>
<level> ::= from 0.1 to 9.3 in NR3 format
When the :MTESt:RMODe command is set to SIGMa, the
:MTESt:RMODe:SIGMa command sets the test sigma level to which a mask
test runs. Test sigma is the best achievable process sigma, assuming no
failures. (Process sigma is calculated using the number of failures per
test.) The test sigma level indirectly specifies the number of waveforms
that must be tested (in order to reach the sigma level).
Query Syntax
:MTESt:RMODe:SIGMa?
The :MTESt:RMODe:SIGMa? query returns the current Sigma level setting.
Return Format
<level><NL>
<level> ::= from 0.1 to 9.3 in NR3 format
See Also "Introduction to :MTESt Commands" on page 320
":MTESt:RMODe" on page 337
Example Code "Example Code" on page 320

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals