Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 451
:TRIGger:GLITch:RANGe
(see page 658)
Command Syntax
:TRIGger:GLITch:RANGe <less_than_time>[suffix],
<greater_than_time>[suffix]
<less_than_time> ::= (15 ns - 10 seconds) in NR3 format
<greater_than_time> ::= (10 ns - 9.99 seconds) in NR3 format
[suffix] ::= {s | ms | us | ns | ps}
The :TRIGger:GLITch:RANGe command sets the pulse width duration for
the selected :TRIGger:GLITch:SOURce. You can enter the parameters in any
order — the smaller value becomes the <greater_than_time> and the larger
value becomes the <less_than_time>.
Query Syntax
:TRIGger:GLITch:RANGe?
The :TRIGger:GLITch:RANGe? query returns the pulse width duration time
for :TRIGger:GLITch:SOURce.
Return Format
<less_than_time>,<greater_than_time><NL>
See Also • "Introduction to :TRIGger Commands" on page 411
• ":TRIGger:GLITch:SOURce" on page 452
• ":TRIGger:GLITch:QUALifier" on page 450
• ":TRIGger:MODE" on page 417