Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 441
:TRIGger[:EDGE]:LEVel
(see page 658)
Command Syntax
:TRIGger[:EDGE]:LEVel <level>
<level> ::= <level>[,<source>]
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
for internal triggers
<level> ::= ±(external range setting) in NR3 format
for external triggers
<source> ::= {CHANnel<n> | EXTernal}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger[:EDGE]:LEVel command sets the trigger level voltage for the
active trigger source.
Query Syntax
:TRIGger[:EDGE]:LEVel? [<source>]
The :TRIGger[:EDGE]:LEVel? query returns the trigger level of the current
trigger source.
Return Format
<level><NL>
See Also • "Introduction to :TRIGger Commands" on page 411
• ":TRIGger[:EDGE]:SOURce" on page 444
• ":EXTernal:RANGe" on page 226
NOTE
If the optional source is specified and is not the active source, the level on the active source
is not affected and the active source is not changed.