Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 211
:DISPlay:DATA
(see page 658)
Command Syntax
:DISPlay:DATA [<format>][,][<area>][,][<palette>]<display data>
<format> ::= {TIFF}
<area> ::= {GRATicule}
<palette> ::= {MONochrome}
<display data> ::= binary block data in IEEE-488.2 # format.
The :DISPlay:DATA command writes trace memory data (a display bitmap)
to the display or to one of the trace memories in the instrument.
If a data format or area is specified, the :DISPlay:DATA command
transfers the data directly to the display. If neither the data format nor
the area is specified, the command transfers data to the trace memory
specified by the :DISPlay:SOURce command. Available trace memories are
PMEMory0- 9 and these memories correspond to the INTERN_0- 9 files in
the front panel Save/Recall menu.
Graticule data is a low resolution bitmap of the graticule area in TIFF
format. This is the same data saved using the front panel Save/Recall
menu or the *SAV (Save) command.
Query Syntax
:DISPlay:DATA? [<format>][,] [<area>][,] [<palette>]
<format> ::= {TIFF | BMP | BMP8bit | PNG}
<area> ::= {GRATicule | SCReen}
<palette> ::= {MONochrome | GRAYscale | COLor}
The :DISPlay:DATA? query reads display data from the screen or from one
of the trace memories in the instrument. The format for the data
transmission is the # format defined in the IEEE 488.2 specification.
If a data format or area is specified, the :DISPlay:DATA query transfers
the data directly from the display. If neither the data format nor the area
is specified, the query transfers data from the trace memory specified by
the :DISPlay:SOURce command.
Screen data is the full display and is high resolution in grayscale or color.
The :HARDcopy:INKSaver setting also affects the screen data. It may be
read from the instrument in 24-bit bmp, 8-bit bmp, or 24-bit png format.
This data cannot be sent back to the instrument.
Graticule data is a low resolution bitmap of the graticule area in TIFF
format. You can get this data and send it back to the oscilloscope.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals