Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 313
:MEASure:VRMS
(see page 658)
Command Syntax
:MEASure:VRMS [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VRMS command installs a screen measurement and starts a
dc RMS value measurement. If the optional source parameter is specified,
the current source is modified.
Query Syntax
:MEASure:VRMS? [<source>]
The :MEASure:VRMS? query measures and outputs the dc RMS value of
the selected waveform. The dc RMS value is measured on an integral
number of periods of the displayed signal. If at least three edges are not
present, the oscilloscope computes the RMS value on all displayed data
points.
Return Format
<value><NL>
<value> ::= calculated dc RMS value in NR3 format
See Also "Introduction to :MEASure Commands" on page 272
":MEASure:SOURce" on page 297
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals