EasyManua.ls Logo

Agilent Technologies InfiniiVision 5000 Series

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
308 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:MEASure:VBASe
(see page 658)
Command Syntax
:MEASure:VBASe [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VBASe command installs a screen measurement and starts a
waveform base value measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:VBASe? [<source>]
The :MEASure:VBASe? query returns the vertical value at the base of the
waveform. The base value of a pulse is normally not the same as the
minimum value.
Return Format
<base_voltage><NL>
<base_voltage> ::= value at the base of the selected waveform in
NR3 format
See Also "Introduction to :MEASure Commands" on page 272
":MEASure:SOURce" on page 297
":MEASure:VTOP" on page 315
":MEASure:VAMPlitude" on page 306
":MEASure:VMIN" on page 310
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Related product manuals