EasyManuals Logo

Agilent Technologies InfiniiVision 5000 Series Programmer's Guide

Agilent Technologies InfiniiVision 5000 Series
788 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #174 background imageLoading...
Page #174 background image
174 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
Public varQueryResult As Variant
Public strQueryResult As String
Private Declare Sub Sleep Lib "kernel32" (ByVal dwMilliseconds As Long)
Sub Main()
On Error GoTo VisaComError
' Create the VISA COM I/O resource.
Set myMgr = New VisaComLib.ResourceManager
Set myScope = New VisaComLib.FormattedIO488
Set myScope.IO = myMgr.Open("TCPIP0::130.29.70.228::inst0::INSTR")
myScope.IO.Clear ' Clear the interface.
' Turn on segmented memory acquisition mode.
myScope.WriteString ":ACQuire:MODE SEGMented"
myScope.WriteString ":ACQuire:MODE?"
strQueryResult = myScope.ReadString
Debug.Print "Acquisition mode: " + strQueryResult
' Set the number of segments to 50.
myScope.WriteString ":ACQuire:SEGMented:COUNt 50"
myScope.WriteString ":ACQuire:SEGMented:COUNt?"
strQueryResult = myScope.ReadString
Debug.Print "Acquisition memory segments: " + strQueryResult
' If data will be acquired within the IO timeout:
'myScope.IO.Timeout = 10000
'myScope.WriteString ":DIGitize"
'Debug.Print ":DIGitize blocks until all segments acquired."
'myScope.WriteString ":WAVeform:SEGMented:COUNt?"
'varQueryResult = myScope.ReadNumber
' Or, to poll until the desired number of segments acquired:
myScope.WriteString ":SINGle"
Debug.Print ":SINGle does not block until all segments acquired."
Do
Sleep 100 ' Small wait to prevent excessive queries.
myScope.WriteString ":WAVeform:SEGMented:COUNt?"
varQueryResult = myScope.ReadNumber
Loop Until varQueryResult = 50
Debug.Print "Number of segments in acquired data: " _
+ FormatNumber(varQueryResult)
Dim lngSegments As Long
lngSegments = varQueryResult
' For each segment:
Dim dblTimeTag As Double
Dim lngI As Long
For lngI = lngSegments To 1 Step -1
' Set the segmented memory index.
myScope.WriteString ":ACQuire:SEGMented:INDex " + CStr(lngI)

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?