Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 413
Use :TRIGger? to query setup information for the TRIGger subsystem.
Return Format
The return format for the TRIGger? query varies depending on the current
mode. The following is a sample response from the :TRIGger? query. In
this case, the query was issued following a *RST command.
:TRIG:MODE EDGE;SWE AUTO;NREJ 0;HFR 0;HOLD +60.0000000000000E-09;
:TRIG:EDGE:SOUR CHAN1;LEV +0.00000E+00;SLOP POS;REJ OFF;COUP DC