722 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
12 Programming Examples
* --------------------------------------------------------------
* In this example we will do the following:
* - Save the system setup to a file for restoration at a later
* time.
* - Save the oscilloscope display to a file which can be
* printed.
* - Make single channel measurements.
*/
private static void Analyze()
{
byte[] ResultsArray; // Results array.
int nLength; // Number of bytes returned from instrument.
/* SAVE_SYSTEM_SETUP - The :SYSTem:SETup? query returns a
* program message that contains the current state of the
* instrument. Its format is a definite-length binary block,
* for example,
* #800002204<setup string><NL>
* where the setup string is 2204 bytes in length.
*/
Console.WriteLine("Saving oscilloscope setup to " +
"c:\\scope\\config\\setup.dat");
if (File.Exists("c:\\scope\\config\\setup.dat"))
File.Delete("c:\\scope\\config\\setup.dat");
// Query and read setup string.
nLength = oscp.DoQueryIEEEBlock(":SYSTem:SETup?",
out ResultsArray);
Console.WriteLine("Read oscilloscope setup ({0} bytes).",
nLength);
// Write setup string to file.
File.WriteAllBytes("c:\\scope\\config\\setup.dat",
ResultsArray);
Console.WriteLine("Wrote setup string ({0} bytes) to file.",
nLength);
/* RESTORE_SYSTEM_SETUP - Uploads a previously saved setup
* string to the oscilloscope.
*/
byte[] DataArray;
int nBytesWritten;
// Read setup string from file.
DataArray = File.ReadAllBytes("c:\\scope\\config\\setup.dat");
Console.WriteLine("Read setup string ({0} bytes) from file.",
DataArray.Length);
// Restore setup string.
nBytesWritten = oscp.DoCommandIEEEBlock(":SYSTem:SETup",
DataArray);
Console.WriteLine("Restored setup string ({0} bytes).",
nBytesWritten);
/* IMAGE_TRANSFER - In this example, we query for the screen
* data with the ":DISPLAY:DATA?" query. The .png format
* data is saved to a file in the local file system.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals