98 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
n/a *OPT? (see page 109) <return_value> ::= 0,0,<license
info>
<license info> ::= <All field>,
<reserved>, <reserved>,
<reserved>, <reserved>,
<reserved>, <Low Speed Serial>,
<Automotive Serial>, <reserved>,
<Secure>, <reserved>,
<reserved>, <reserved>,
<reserved>,
<RS-232/UART Serial>, <reserved>
<All field> ::= {0 | All}
<reserved> ::= 0
<Low Speed Serial> ::= {0 | LSS}
<Automotive Serial> ::= {0 | AMS}
<Secure> ::= {0 | SEC}
<RS-232/UART Serial> ::= {0 |
232}
*RCL <value> (see
page 110)
n/a <value> ::= {0 | 1 | 2 | 3 | 4 |
5 | 6 | 7 | 8 | 9}
*RST (see page 111) n/a See *RST (Reset) (see page 111)
*SAV <value> (see
page 114)
n/a <value> ::= {0 | 1 | 2 | 3 | 4 |
5 | 6 | 7 | 8 | 9}
*SRE <mask> (see
page 115)
*SRE? (see page 116) <mask> ::= sum of all bits that
are set, 0 to 255; an integer in
NR1 format. <mask> ::= following
values:
Bit Weight Name Enables
--- ------ ---- ----------
7 128 OPER Operation Status Reg
6 64 ---- (Not used.)
5 32 ESB Event Status Bit
4 16 MAV Message Available
3 8 ---- (Not used.)
2 4 MSG Message
1 2 USR User
0 1 TRG Trigger
Table 29 Common (*) Commands Summary (continued)