Description Setting
Size calling method
Local Southern (default)—Determines the fragment sizes using the reciprocal
relationship between fragment length and electrophoretic mobility.
Global Southern—Compensates for standard fragments with anomalous
electrophoretic mobility (similar to least squares methods).
Second Order Least Squares—Uses regression analysis to build a best-fit size
calling curve.
Third Order Least Squares—Uses regression analysis to build a best-fit size
calling curve.
Cubic Spline—Forces the sizing curve through all the known points of the
selected size standard.
Chapter 11 Manage library resources on the instrument
Analysis settings library
11
296
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide