Spectral calibration with Allow borrowing disabled
Table 9 Overview of spectral calibration options with Allow borrowing disabled
Option Action
Allow borrowing
disabled
Auto Retry disabled
All capillaries must pass (must meet the Spectral QV and Condition
number limits set internally in the software).
One run is performed (see Injection 1 below).
If the limits are not met, the calibration fails.
This option is not illustrated in the table below.
Allow borrowing
disabled
Aut
o Retry enabled
Same as above, except up to 3 runs are performed (see Injections 1 through 3
below).
Table 10 Details of spectral calibration injections with Allow borrowing disabled and Auto Try
enabled
Injection Process
Injection 1
The softwar
e evaluates the Spectral QV and Condition number of all capillaries.
If all capillaries pass, the calibration is complete, and injections 2 and 3 are not
performed.
If any capillaries fail and Auto Retry is disabled, the calibration fails.
If any capillaries fail and Auto Retry is enabled, injection 2 is performed.
Injection 2 (listed as
re-injection in
Run queue)
The softwar
e evaluates the Spectral QV and Condition number for each capillary
across injections 1 and 2 and uses the information from the capillary with the
highest Spectral QV.
If all capillaries now pass, the calibration is complete and injection 3 is not
performed.
If the same capillary fails in both injection 1 and 2, injection 3 is performed.
Injection 3 (listed as
re-injection in
Run queue)
The softwar
e evaluates the Spectral QV and Condition number for each capillary
across injections 1, 2, and 3 and uses the information from the capillary with the
highest Spectral QV.
If all capillaries now pass, the calibration passes.
If the same capillary fails in injection 1, 2, or 3 and the calibration fails.
Chapter 12 Run calibrations and install checks
Run a spectral calibration
12
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
323