Spectral calibration with borrowing enabled
Table 7 Overview of spectral calibration options with Allow borrowing enabled
Option Action
Allow borrowing enabled
Auto Retry disabled
(default)
All capillaries must pass (must meet the Spectral QV and Condition
number limits set internally in the software) and meet the borrowing limits
listed below.
8-capillary instruments—One adjacent-capillary borrowing event is
allowed
24-capillary instruments—Up to two adjacent-capillary borrowing
events are allowed.
One run is performed (see Injection 1 below).
If the limits are not met, the calibration fails.
This option is not illustrated in the table below.
Allow borrowing enabled
Auto Retry enabled
Same as above, except up to 3 runs are performed if injections 1 and 2 fail (see
Injections 1 through 3 below).
Table 8 Details of spectral calibration injections with Allow borrowing enabled and Auto Try enabled
Injection Process
Injection 1
The softwar
e evaluates the Spectral QV and Condition number of all capillaries.
If all capillaries pass, the calibration is complete, and injections 2 and 3 are not
performed.
If any capillaries fail, the software borrows from an adjacent capillary.
If, after borrowing, >1 (8-capillary instruments) or > 2 capillaries (24-capillary
instruments) fail, injection 2 is performed.
Injection 2 (listed as
re-injection in
Run queue)
The softwar
e evaluates the Spectral QV and Condition number results between
adjacent capillaries in injection 2 and for each capillary across injections 1 and 2.
It uses the result with the highest Spectral QV for each capillary.
If all capillaries pass, the calibration is complete and injection 3 is not performed.
If, after borrowing, >1 (8-capillary instruments) or > 2 capillaries (24-capillary
instruments) from injection 1 or 2 do not pass, injection 3 is performed.
Injection 3 (listed as
re-injection in
Run queue)
The softwar
e evaluates the Spectral QV and Condition number results between
adjacent capillaries in injection 3 and for each capillary across injections 1, 2,
and 3. It uses the information with the highest Spectral QV for each capillary.
If all capillaries now pass, the calibration passes.
If after borrowing, >1 (8-capillary instruments) or > 2 capillaries (24-capillary
instruments) from injection 1, 2, or 3 do not pass, the calibration fails.
Chapter 12 Run calibrations and install checks
Run a spectral calibration
12
322
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide