Prepare the install run standard
For the instructions to prepare the standard, see the product information sheet. To find the product
information sheet, go to www.thermofisher.com, search for the product, then scroll to the bottom of
the product page.
Prepare the appropriate standard according to the procedure in the product information sheet.
Note: If you use the BigDye
™
Direct Cycle Sequencing Kit, perform the install run with BigDye
™
v3.1 sequencing standards. For more information, see “Using BigDye
™
Direct Cycle Sequencing Kit
chemistry” on page 312.
Install check type Reagent
Sequencing
•
Sequencing Standards, BigDye
™
Terminator v1.1 (Cat. No. 4404314)
•
Sequencing Standards, BigDye
™
Terminator v3.1 (Cat. No. 4404312)
Sequencing for
MicroSEQ
™
ID
applications
Sequencing Standards, BigDye
™
Terminator v1.1 (Cat. No. 4404314)
Fragment DS‑33 GeneScan
™
Installation Standards with GeneScan
™
600 LIZ
™
Size
Standard v2.0 (Cat. No. 4376911)
Chapter 12 Run calibrations and install checks
Perform an install run
12
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
331