EasyManua.ls Logo

Thermo Scientific appliedbiosystems SeqStudio Flex Series - Page 368

Thermo Scientific appliedbiosystems SeqStudio Flex Series
574 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.
Check that the seal is intact. Do not use if the buer level is too low or the seal has been
compromised. A fill tolerance of ±0.5 mm is acceptable.
1
1
Fill line
8.
T
ilt the CBC back and forth gently and carefully to ensure that the buer is evenly distributed
across the top of the baes. If you do not tilt the CBC back and forth, the buer sticks to the
baes because of surface tension.
9.
Verify that the buer is at or above the fill line.
10.
When ready to install the CBC, place the container on a flat surface (such as a lab bench) and peel
o the seal.
IMPORTANT! Mak
e sure that all pieces of the seal are removed. Remnants of the seal can block
holes in the CBC.
11.
W
ipe o any buer on top of the CBC with a lint-free tissue. Ensure that the top of the container is
dry. Moisture can cause arcing and termination of a run.
12.
Place the appropriate septum on each side of the CBC:
a.
Align the buer septum (the part that is symmetrical) over the 24 holes of the CBC.
b.
Push the septum lightly into the holes to start, then push firmly to seat it.
c.
Align the capillary washing septum over the other chamber of the CBC.
Position the rounded starter tab of the septum over the larger orientation hole on the washing
chamber then press down. Press down on the remaining holes in the septum.
Chapter 13 Maintain the Instrument
Install buers (wizar
d)
13
368
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide

Table of Contents

Related product manuals