f.
Tap Next to continue.
Change Polymer Type or Shut Down Instrument—Insert array port plug
In this procedure, you install a plug in the array port. This plug allows pump washing without a capillary
array installed.
1.
Obtain the array port plug.
1
Figure 46 Obtain the array port plug
1
Obtain the array port plug in holder on pump block above the ABC
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
13
398
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide