Table 17 Sequencing analysis run modules (continued)
Run module
type
Run module name
Configuration 23 hours throughput
[1]
Perform
- ance
Cap.
length
(cm)
Polymer
type
Run
time
(min)
(8cap.) (24-cap.)
Contig.
Read
Length
(CRL)
[2]
Rapid
sequencing
BigDye
XTerminator
BDxRapidSeq50_POP6
BDxRapidSeq50_POP6xl
50 POP-6
≤68 ≥160 ≥480 ≥450
BDxRapidSeq50_POP7
BDxRapidSeq50_POP7xl
50 POP-7
≤40 ≥280 ≥840 ≥500
Rapid
sequencing
BigDye
XTerminator
BDxRapidSeq36_POP4
BDxRapidSeq36_POP4xl
36 POP-4
≤45 ≥240 ≥720 ≥400
BDxRapidSeq36_POP6
BDxRapidSeq36_POP6xl
36 POP-6
≤70 ≥156 ≥468 ≥600
BDxRapidSeq36_POP7
BDxRapidSeq36_POP7xl
36 POP-7
≤33 ≥328 ≥984 ≥600
Fast
sequencing
FastSeq50_POP6
FastSeq50_POP6xl
50 POP-6
≤90 ≥122 ≥368 ≥600
FastSeq50_POP7
FastSeq50_POP7xl
50 POP-7
≤65 ≥168 ≥504 ≥700
Fast
sequencing
FastSeq36_POP7
FastSeq36_POP7xl
36 POP-7
≤60 ≥184 ≥552 ≥750
Fast
sequencing
BigDye
XTerminator
BDxFastSeq50_POP6
BDxFastSeq50_POP6xl
50 POP-6
≤90 ≥122 ≥368 ≥600
BDxFastSeq50_POP7
BDxFastSeq50_POP7xl
50 POP-7
≤65 ≥168 ≥504 ≥700
Fast
sequencing
BigDye
XTerminator
BDxFastSeq36_POP7
BDxFastSeq36_POP7xl
36 POP-7
≤60 ≥184 ≥552 ≥750
Standard
sequencing
StdSeq50_POP6
StdSeq50_POP6xl
50 POP-6
≤135 ≥80 ≥240 ≥600
StdSeq50_POP7
StdSeq50_POP7xl
50 POP-7
≤125 ≥88 ≥264 ≥850
Appendix B Run modules and dye sets
Run modules, read lengths, size ranges, and run times
B
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide