EasyManua.ls Logo

Thermo Scientific appliedbiosystems SeqStudio Flex Series - Page 529

Thermo Scientific appliedbiosystems SeqStudio Flex Series
574 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 18 Fragment analysis run modules
Run
module
type
Run module
name
Configuration 23 hours throughput
[1]
Performance
Cap.
length
(cm)
Pol. type
Run
time
(min)
(8-
cap.)
(24-
cap.)
Range
[2]
Sizing precision
[3]
50–
400bp
401–
600bp
601–
1,200
bp
Fragment
analysis
FragmentAnalysi
s50_POP7
FragmentAnalysi
s50_POP7xl
50 POP-7
≤45 ≥280 ≥840
[4]
≥40 to
≤520
<0.15 <0.30 NA
[5]
FragmentAnalysi
s50_POP6
FragmentAnalysi
s50_POP6xl
50 POP-6
≤100 ≥112 ≥336 ≥20 to
≤550
<0.15 <0.30 NA
[5]
FragmentAnalysi
s36_POP4
FragmentAnalysi
s36_POP4xl
36 POP-4
≤35 ≥312 ≥936 ≥60 to
≤400
<0.15 NA NA
[5]
FragmentAnalysi
s36_POP7
FragmentAnalysi
s36_POP7xl
36 POP-7
≤30 ≥368 ≥1,104 ≥60 to
≤400
<0.15 NA NA
[5]
Fragment
analysis
FragAnalysis36_
POP6
FragAnalysis36_
POP6xl
36 POP-6
≤66 ≥160 ≥480 ≥60 to
≤500
<0.15 <0.30 NA
[5]
Appendix B Run modules and dye sets
Run modules, read lengths, size ranges, and run times
B
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
529

Table of Contents

Related product manuals