4.
In the next Edit Injection Properties screen, specify the settings for the injection.
Specify settings by doing either of the following:
•
In the Injection pr
otocol field, select a protocol.
An injection protocol contains the elements listed below. When you select an injection
protocol, these elements are automatically selected. However, you can tap any of the fields
below the Injection protocol field to select a dierent element for the plate file as described
below. Swipe up to display the rest of the screen.
•
Select individual elements in the following fields. Selecting an individual element overrides an
injection protocol selection and clears the Injection protocol field.
–
Dye set
–
Run module
–
Analysis settings
–
Size standard (fragment analysis only)
Note: If you use the BigDye
™
Direct Cycle Sequencing Kit, select Z_BigDye Direct for the
dye set. For more information, see “Using BigDye
™
Direct Cycle Sequencing Kit chemistry” on
page 312.
For information on creating injection protocols and other run settings, see Chapter 11, “Manage
libr
ary resources on the instrument”.
Chapter 4 Cr
eate a plate file on the instrument
Enter injection properties
4
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
91