EasyManuals Logo

Tektronix MSO58 User Manual

Tektronix MSO58
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #122 background imageLoading...
Page #122 background image
Commands listed in alphabetical order
Returns
The query respo
nse is returned in high precision NR3 format (up to 12 digits with
more than 4 trailing 0 digits after the decimal point is omitted).
Examples
AFG:PERIOD 1
sets the AFG period value to 1 second.
AFG:PERIOD? might return :AFG:PERIOD 3.2000E-6 indicating the AFG p eriod
is set to 3.2 μs.
AFG:PULse:WIDth
Sets (or queries) the AFG pulse width, in seconds.
Conditions
Requires option AFG.
Group
AFG
Syntax
AFG:PULse:WIDth <NR3>
AFG:PULse:WIDth?
Arguments
NR3 is the oating point number that represents the pulse width, in seconds.
Examples
AFG:PULSE:WIDTH 100.0E-6 sets the AFG pulse width to 100 microseconds.
AFG:PULSE:WIDTH? might return :AFG:PULSE:WIDTH 1.0000E-6 indicating
the pulse width is set to 1 μs.
AFG:RAMP:SYMmetry
Sets (or queries) the AFG ramp symmetry in percent. Minimum is 0.0%,
maximum is 100.0% and increment is 0.10%.
Conditions
Requires option AFG.
Group
AFG
Syntax
AFG:RAMP:SYMmetry <NR3
>
AFG:RAMP:SYMmetry?
Arguments
Floating point number that represents the A FG ramp symmetry, as a percentage.
2-106 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals