Commands listed in alphabetical order
Returns
The query respo
nse is returned in high precision NR3 format (up to 12 digits with
more than 4 trailing 0 digits after the decimal point is omitted).
Examples
AFG:PERIOD 1
sets the AFG period value to 1 second.
AFG:PERIOD? might return :AFG:PERIOD 3.2000E-6 indicating the AFG p eriod
is set to 3.2 μs.
AFG:PULse:WIDth
Sets (or queries) the AFG pulse width, in seconds.
Conditions
Requires option AFG.
Group
AFG
Syntax
AFG:PULse:WIDth <NR3>
AFG:PULse:WIDth?
Arguments
NR3 is the oating point number that represents the pulse width, in seconds.
Examples
AFG:PULSE:WIDTH 100.0E-6 sets the AFG pulse width to 100 microseconds.
AFG:PULSE:WIDTH? might return :AFG:PULSE:WIDTH 1.0000E-6 indicating
the pulse width is set to 1 μs.
AFG:RAMP:SYMmetry
Sets (or queries) the AFG ramp symmetry in percent. Minimum is 0.0%,
maximum is 100.0% and increment is 0.10%.
Conditions
Requires option AFG.
Group
AFG
Syntax
AFG:RAMP:SYMmetry <NR3
>
AFG:RAMP:SYMmetry?
Arguments
Floating point number that represents the A FG ramp symmetry, as a percentage.
2-106 MSO54, MSO56, MSO58, MSO58LP Programmer