Commands listed in alphabetical order
MEASUrement:
REFLevels:ABSolute:FALLLow
This command sets or que ries the value used as the low reference l evel of the
falling edge when the measurement's ref level method is set to absolute.
Group
Measurement
Syntax
MEASUrement:REFLevels:ABSo lute:FALLLow <NR3>
MEASUrement:REFLevels:ABSo lute:FALLLow?
Arguments
<NR3> is the value used as the low reference level of the falling edge.
Examples
MEASUrement:REFLevels:ABSo lute:FALLLow -1.5 sets the low reference
level to -1.5 V.
MEASUrement:REFLevels:ABSo lute:FALLLow? might return
:MEASUREMENT:REFLEVELS:ABS OLUTE:FALLLOW -1.0000 indicating the
low reference level is -1.0 V.
MEASUrement:REFLevels:ABSolute:FALLMid
This command sets or queries the value used as the mid reference level of the
falling edge when the measurement's ref level method is set to absolute.
Group
Measurement
Syntax
MEASUrement:REFLevels:ABSo lute:FALLMid <NR3>
MEASUrement:REFLevels:ABSo lute:FALLMid?
Arguments
<NR3> is the value used as the mid reference level of the falling edge.
Examples
MEASUrement:REFLevels:ABSo lute:FALLMid 10.0E-3 sets the mid
reference level to 10.0 mV.
MEASUrement:REFLevels:ABSo lute:FALLMid? might return
:MEASUREMENT:REFLEVELS:ABS OLUTE:FALLMID 0.0E+0 indicating the mid
reference level is 0.0 V.