Commands listed in alphabetical order
Arguments
1|ONselects th
e gating settings as Global.
0 | OFF selects the gating settings as Local.
Examples
POWer:POWer1:GATing:GLOBal 1 selects t he gating settings as Global for the
power measurement badge 1.
POWer:POWer<x>:HARMONICS:CLASs
This command sets or queries the class type for the harmonics measurement
in the specied power measurement badge. The power measurement badge
is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:HARMONICS:C LASs {CLASSA | CLASSB | CLASSC |
CLASSD}
POWer:POWer<x>:HARMONICS:C LASs?
Related Commands
POWer:POWer<x>:HARMONICS:STANDard
Examples
POWer:POWer1:HARMONICS:CLA Ss CLASSB sets the class type for harmonics
measurement as CLA SSB.
POWer:POWer<x>:HARMONICS:CMEThod
This command sets or queries the fundamental current method for the harmonics
measurement in the specied power measurement badge. The power measurement
badge is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:HARMONICS:C METhod {RATed | MEASured}
POWer:POWer<x>:HARMONICS:C METhod?
MSO54, MSO56, MSO58, MSO58LP Programmer 2-537