Commands listed in alphabetical order
POWer:POWer<
x>:SWITCHINGLOSS:RDSOn
This command sets or queries the RDS(on) value for switching loss measurement
in the specied power measurement badge. The power measurement badge
is specied b
yx.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:SWITCHINGL OSS:RDSOn <NR1>
POWer:POWer<x>:SWITCHINGL OSS:RDSOn?
Related Commands
POWer:POWer<x>:SWITCHINGLOSS:DEVICEType
Arguments
<NR1> ranges from 0 to 100
Examples
POWer:POWer1:SWITCHINGLOS S:RDSOn 2 sets the switching loss RDSOn
value as 2 for switching loss measurement of the power measurement badge
Power 1.
POWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType
This command sets or queries the conguration type for the switching loss
measurement in the specied power m easurement badge. The power measurement
badge is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:SWITCHINGL OSS:SWLCONFIGType {SMPS | PFC |
FLYBACK}
POWer:POWer<x>:SWITCHINGL OSS:SWLCONFIGType?
Arguments
SMPS: Select this option in case of signals without noise and ringing. The V
g
source is not required. Select V
g
souce (Source 3), in case of noisy signal.
PFC: Select this option when input DUT signals are from Power Factor Correction
Circuit. For this case, V
g
source is mandatory.